BIOSENSE WEBSTER INC OCTARAY MAPPING CATHETER; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY
|
Back to Search Results |
|
| Catalog Number |
D160904 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure and the octaray mapping catheter was not able to flush after being used at the beginning of the case, and right at mid-case, when it was ¿required to back in¿.The catheter was replaced, and the problem was resolved, and the case continued.No adverse patient consequence was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure and the octaray mapping catheter was not able to flush after being used at the beginning of the case, and right at mid-case, when it was ¿required to back in¿.The catheter was replaced, and the problem was resolved, and the case continued.No adverse patient consequence was reported.Device investigation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Visual inspection and irrigation test of the returned device were performed following j&j medtech procedures.Visual analysis revealed no damage or anomalies on the device.An irrigation test was performed, and the device was irrigated correctly.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device number lot 31458101l and no internal action related to the complaint was found during the review.The irrigation issue reported by the customer could not be replicated during the product investigation.Other issues or circumstances may have occurred during the usage of the device that compromised its performance.The instructions for use contain the following recommendations: flush the catheter with heparinized saline prior to insertion into the body.As part of johnson & johnson medtech¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|