• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CENTURION VISION SYSTEM, HANDPIECE TIP; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CENTURION VISION SYSTEM, HANDPIECE TIP; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065790023
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
A customer via government agency reported needle was broken during cataract surgery (with intraocular lens implant).The surgery was completed by replacing needle with a new one.There was no patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in h.6.And h.11.A sample was not received at the investigation site for evaluation.Therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, HANDPIECE TIP
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21205066
Report Number2523835-2025-00067
Device Sequence Number11748146
Product Code HQC
UDI-Device Identifier00380657900237
UDI-Public(01)00380657900237(17)280331(11)230427(10)15KCUC
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K121555
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065790023
Device Lot Number15KCUC
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/06/2025
Supplement Date Received by Manufacturer04/02/2025
Initial Report FDA Received Date01/22/2025
Supplement Report FDA Received Date04/25/2025
Date Device Manufactured04/27/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexFemale
-
-