| Catalog Number |
0170SI16 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that foley coude catheter inserted during night shift from (b)(6)2024.Foley catheter was leaking significantly during physical therapy.Nurse deflated balloon; 9 milliliters of normal saline came out.Nurse reinflated balloon and gently pulled back to confirm balloon inflation.The catheter came all the way out at that point and the balloon was not inflated.It appeared the balloon had a leak or tear during the re-inflation attempt.Unfortunately, they were not able to obtain a lot number as the packaging was disposed of during the night shift trash takeout.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.No actions can be taken at this time since a root cause was not identified.A dhr could not be performed since no lot number was provided.The instructions for use were found adequate and state the following: 1.Occlude drainage tubing a minimum of 3 inches below the sampling port by kinking the tubing until urine is visible under the access site.2.Swab surface of bard ez-lok sampling port with antiseptic wipe.(fig.2) 3.Using aseptic technique, position the syringe in the center of the sampling port.Press the syringe firmly and twist gently to access the sampling port.(fig.3) 4.Slowly aspirate urine sample into syringe and remove syringe from sample port.5.Unkink tubing if necessary and transfer urine specimen into specimen cup or follow hospital protocol.Discard syringe according to hospital protocol.6.Follow established hospital protocol for specimen labeling and transport to lab.Sterile unless package is opened or damaged, except for any individually packaged components within the tray which are not labeled as sterile.These components are not terminally sterilized.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not resterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Proper techniques for urinary catheter insertion perform hand hygiene immediately before and after insertion insert urinary catheters using aseptic technique and sterile equipment use the smallest foley catheter possible, consistent with good drainage document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in patient record proper techniques for urinary catheter maintenance secure the foley catheter, use the statlock foley stabilization device if provided maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions maintain unobstructed urine flow and keep the catheter and collection tube free from kinking keep the collection bag below the level of the bladder or hips at all times empty the collection bag regularly (e.G., prior to transport) using a separate, clean collection container for each patient routine hygiene (e.G., cleansing of the meatal surface during daily bathing or showering) is appropriate leave foley catheter in place only as long as needed udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that foley coude catheter inserted during night shift from (b)(6) 2024 to (b)(6) 2024.Foley catheter was leaking significantly during physical therapy.Nurse deflated balloon, 9 milliliter of normal saline came out.Nurse reinflated balloon and gently pulled back to confirm balloon inflation.The catheter came all the way out at that point and the balloon was not inflated.It appeared the balloon had a leak or tear during the re-inflation attempt.Unfortunately, they were not able to obtain a lot number as the packaging was disposed of during the night shift trash takeout.
|
| |
|
Event or Problem Description
|
|
It was reported that foley coude catheter inserted during night shift from (b)(6) 2024.Foley catheter was leaking significantly during physical therapy.Nurse deflated balloon, 9 milliliter of normal saline came out.Nurse reinflated balloon and gently pulled back to confirm balloon inflation.The catheter came all the way out at that point and the balloon was not inflated.It appeared the balloon had a leak or tear during the re-inflation attempt.Unfortunately, they were not able to obtain a lot number as the packaging was disposed of during the night shift trash takeout.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is unconfirmed.Visual: received one (1) used lubri-sil all silicone coude tip foley catheter attached to the urine meter bag with label without original packaging.The catheter balloon was inflated with 10ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and the catheter rested with no leaks noted.The balloon passively deflated in under one (1) minute with no cuffing or leaks noted, returning 10ml of solution.This is within specification per inspection procedure ip7603444, revision 0, which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.As the reported event is unconfirmed, a risk and labeling review is not required.A review of the device history record was not necessary due to the reported event being unconfirmed.Based on the investigation results no additional action is required at this time.Correction: d,e,h udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|