• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS LIPC; LIPASE-ESTERASE, ENZYMATIC, PHOTOMETRIC, LIPASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS LIPC; LIPASE-ESTERASE, ENZYMATIC, PHOTOMETRIC, LIPASE Back to Search Results
Catalog Number 03183742122
Medical Device Problem Code Device Difficult to Setup or Prepare (1487)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The cobas c502 module serial number was (b)(6).Qc was within the ranges.The alarm trace provided from (b)(6) 2024 showed abnormal aspiration alarms 66 times and abnormal cell blank alarms 31 times.The customer exchanged the photometer lamp and cuvettes regularly according to specifications and cleaned the sample and reagent needles daily.The gear pump had not been replaced and the field service engineer confirmed that the gear pump pressure more than 300kpa.The investigation is ongoing.
 
Event or Problem Description
We received an allegation about questionable results for 1 patient serum/plasma samples not matching the patient's clinical diagnosis when tested with amyl gen.2 (amyl2) assay and lipase colorimetric assay (lipc) assay on a cobas 8000 cobas c502 module.This medwatch will apply to the lipc assay.Please refer to the medwatch with a1.Patient identifier (b)(6) for information related to the amyl2 assay.On (b)(6) 2024, the 1st sample (serum/plasma sample) was tested resulting in: amyl2 result: 576 u/l.Lipc result: 1692.5 u/l.On (b)(6) 2024, the physician requested a 2nd sample to be drawn and tested resulting in: amyl2 result: 338 u/l (urine sample).Amyl2 result: 104 u/l (serum sample).Lipc result: 47.8 u/l (serum sample).On (b)(6) 2024, the physicians questioned the results and requested a 3rd blood sample to be drawn and tested resulting in: amyl2 result: 195 u/l.Lipc result: 259.3 u/l.On (b)(6) 2024, the original sample was repeated resulting in: amyl2 result: 577 u/l.Lipc result: 1644.4 u/l.The results did not match the patient's clinical diagnosis and expectations.
 
Additional Manufacturer Narrative
A general reagent problem can be excluded because calibration and qc were acceptable.Based on the high amount of sample-related alarms and the fact that the pre-analytics were not handled according to the tube instruction for use (target volume vs draw volume and an incorrect centrifugation speed), the cause of the event was consistent with a preanalytic issue at the customer site.Preanalytics are within the customer's responsibility.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIPC
Common Device Name
LIPASE-ESTERASE, ENZYMATIC, PHOTOMETRIC, LIPASE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250-0457
MDR Report Key21209354
Report Number1823260-2025-00199
Device Sequence Number11715033
Product Code CHI
UDI-Device Identifier04015630918935
UDI-Public04015630918935
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K972250
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Catalogue Number03183742122
Device Lot Number806509
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/27/2024
Supplement Date Received by Manufacturer02/24/2025
Initial Report FDA Received Date01/22/2025
Supplement Report FDA Received Date02/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
Patient RaceAsian
-
-