| Catalog Number |
ENC453712 |
| Medical Device Problem Codes |
Premature Activation (1484); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #:(b)(4).Updated sections on this med watch report: b4, d1, g3, g6, h2, and h11 correction: section d1.Brand name: cerenovus enterprise section 2a.Common device name: intracranial neurovascular stent a supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section e1.Initial reporter phone: (b)(6).Section h3 - the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
As reported by the field, during an endovascular embolization, an eu 4.5x37mm stent 12 mm dw tip intracranial stent (enc453712, 8945944) became impeded in the proximal of a prowler select plus 150/5cm microcatheter (606s255x, 31392027) and could not advance any more.The physician tried to retract the stent, the stent was found detached from the delivery wire in the microcatheter (mc).The doctor removed the microcatheter and stent from the patient and switch new devices to complete the surgery.Additional event information received on 20-jan-2025 indicated that they were able to torque the device.There was no evidence of physical material within the device.The microcatheter did not kink/bent.There was no excessive force used with the devices.The event did not prolog the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Updated sections on this medwatch: b4, b5, g3, g6, h2, h3, h6 and h11.Section b5: additional event information received on 05-feb-2025 confirmed that the type of procedure was an angioplasty.Complaint conclusion: as reported by the field, during an angioplasty procedure, an eu 4.5x37mm stent 12 mm dw tip intracranial stent (enc453712, 8945944) became impeded in the proximal of a prowler select plus 150/5cm microcatheter (606s255x, 31392027) and could not advance any more.The physician tried to retract the stent, the stent was found detached from the delivery wire in the microcatheter (mc).The doctor removed the microcatheter and stent from the patient and switch new devices to complete the surgery.Additional event information received on 20-jan-2025 indicated that they were able to torque the device.There was no evidence of physical material within the device.The microcatheter did not kink/bent.There was no excessive force used with the devices.The event did not prolog the procedure.The device was discarded; therefore, no further investigation can be performed.A device history record (dhr) was performed, and it indicated this product was final inspection tested at lake region medical and was determined to be acceptable.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the device history record review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be determined; however, there are circumstances of the procedure that may have contributed to the reported failure.The instructions for use (ifu) do contain the following recommendations: ¿ do not apply undue force if resistance is encountered at any point during stent manipulation.Withdraw the unit and advance to a new one.¿ if resistance is felt while recapturing the stent, do not continue to recapture the device.Withdraw the infusion catheter slightly to unsheathe the stent (without exceeding the recapture limit), and then attempt to recapture the stent again.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).The device was returned to j&j medtech for further evaluation.A non-sterile eu 4.5x37mm stent 12 mm dw tip intracranial stent was received contained in the decontamination pouch.Upon receiving the device, visual inspection was performed, and the stent was detached from the delivery unit.Microscopic inspection was performed on the stent component.It was observed to be in good condition; there was no structural damage (i.E., no broken struts, no kinks); also, it was noted fully expanded, both ends can be noted as completely flared.The delivery wire was subjected to dimensional analysis and all measurements were found to be within specification, including those specifications that control the attachment and delivery of the stent.Therefore, device failure is not suspected to be a contributing factor.The customer complaint regarding a stent being automatically released was confirmed since the stent was noted as already separated from the delivery system; based on this condition, the issue regarding a stent being impeded in the proximal section of the concomitant microcatheter cannot be evaluated through functional testing.The stent must be inside the introducer tube to perform the functional analysis.Additionally, none of the returned components present damages that suggest that they were forcibly advanced.With the limited information available, a conclusive cause cannot be determined; however, it is possible that clinical and procedural factors, including device manipulation and operator's technique, may have contributed to the reported failure.At this time, there is no evidence to support that the issue reported in the complaint is a result of a defect inherently related to the device.Lake region medical did review the device history records relative to the manufacturing, inspecting and packaging of the lot 8945944.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As part of the j&j medtech quality process, all devices are manufactured, inspected, and released to approved specifications.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no capa activity is required.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) do contain the following recommendations: do not partially deploy the stent from the introducer.Confirm that the delivery wire does not move relative to the introducer during the removal of the enterprise vascular reconstruction device and delivery system from the dispenser hoop.Confirm the tip of the delivery wire is entirely within the introducer.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|