| Catalog Number |
2B12LT |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/30/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent: 1/22/2025 d4: batch # unk.Additional information was requested and the following was obtained: "surgery name: laparoscopic renal section" attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown surgery, after insert into the trocar, noted the device was broken.Another device was used to complete the surgery.There was no patient consequence reported.No additional information could be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/3/2025.D4: batch # c9df1n.Investigation summary the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that the 2b12lt device was returned with the universal seal component disassembled and inside a plastic bag.In addition, the packaging opened was returned along with the instrument.Upon visual evaluation of the universal seal was observed to be not properly welded.No conclusion could be reached regarding what may have caused the reported incident.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|