Catalog Number 163662
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 12/23/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a revision procedure due to dislocation at the site.There is no additional information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).D10: (b)(6) item name ringloc bi-polar 28x54mm lot # 66949684.The device will not be returned for analysis as it¿s location is not known; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event or Problem Description
There is no update to the prior event description provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: updated: b4, d4: lot, d9, g3, g6, h2, h11.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; d4; d9; g3; h2; h3; h4; h6.H6: component code: mechanical (g04) - head.Product returned and evaluated.Visual examination of the returned product identified that the liner and shell were returned assembled with the head disassembled.Scratches were noted on the spherical surface of the head and taper hole were from use.No other damage was noted.The shell outer spherical surface had scratches.Liner had damage indentation/deformation on the rim.No other damage was noted.Measurements taken of the product were found to be conforming to specification.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported products were reviewed for compatibility with no issues noted.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: loose and displaced acetabular cup with subsequent implant dislocation as described.A definitive root cause cannot be determined.This complaint was confirmed based on the radiograph review.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4;g3;h2;h3.The following section was corrected: d4.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
28MM DIA COCR MOD HD +3MM NK
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21211256
Report Number0001825034-2025-00191
Device Sequence Number15457677
Product Code JDG
UDI-Device Identifier00887868308277
UDI-Public(01)00887868308277(17)340820(10)J7815395
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K942479
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number163662
Device Lot NumberJ7815395
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/02/2025
Supplement Date Received by Manufacturer03/13/2025
06/09/2025
05/21/2025
Initial Report FDA Received Date01/22/2025
Supplement Report FDA Received Date03/25/2025
06/11/2025
06/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;