• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. UNIVERSAL ENDOSCOPIC ULTRASOUND CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. UNIVERSAL ENDOSCOPIC ULTRASOUND CENTER Back to Search Results
Model Number EU-ME1
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/31/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that while under sedation for a diagnostic endobronchial ultrasound-guided transbronchial needle aspiration, the ultrasound image quality was poor.The issue occurred mid-procedure causing a surgical prolongation greater than 30 minutes.The issue was resolved after utilization of a new scope and balloon.There were no reports of patient harm.
 
Event or Problem Description
This supplemental report is being submitted to provide additional information from the customer and the results of the legal manufacturer¿s investigation.Updated fields: b5, h2 and h6.Corrections to b1, b2, and h1.This event was initially conservatively reported as a serious injury; however, upon further follow-up, the customer confirmed that the patient did not suffer any serious injury or adverse effects from the prolonged procedure, thus correcting b1, b2, and h1.B1 should not be marked as an adverse event, and b2 should not be marked at all.H1 should also not be marked.The h6 medical device problem reported and other findings would not cause or contribute to a death or a serious injury if it were to recur.The device was not returned to olympus for inspection and the reported failure was not confirmed.A definitive root cause could not be identified due to no findings available.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL ENDOSCOPIC ULTRASOUND CENTER
Common Device Name
UNIVERSAL ENDOSCOPIC ULTRASOUND CENTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
MDR Report Key21211846
Report Number2429304-2025-00041
Device Sequence Number11719281
Product Code IYN
UDI-Device Identifier04953170355639
UDI-Public04953170355639
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/08/2025
Report Date (Section F) 04/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberEU-ME1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date03/14/2025
Device Age12 YR
Event Location Hospital
Date Report to Manufacturer12/31/2024
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/22/2025
Supplement Report FDA Received Date04/08/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BALLOON; SCOPE
Patient SexUnknown
Date Report Sent to FDA04/08/2025
-
-