COVIDIEN SURGICAL V-LOC 90; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC Back to Search Results
Model Number VLOCM0023
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, in an open joint replacement procedure, when closing the wound using the continuous suturing technique, one suture had its needle detached from the thread two minutes into the procedure.Two other sutures had broken threads two minutes into the procedure.Another device was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant products: vlocl0315, vlocl0315 v-loc 180 2-0 grn 30cm gs21, (lot # a3h2127vy); vlocl0315, vlocl0315 v-loc 180 2-0 grn 30cm gs21, (lot # a3l1837vy).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d3(mfr name and street 1), d4(udi), d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.One representative sample was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the suture was broken.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-LOC 90
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21215689
Report Number1219930-2025-00400
Device Sequence Number3517871
Product Code GAM
UDI-Device Identifier10884521131682
UDI-Public(01)10884521131682(17)251231(10)A3A0158VFY(20)01
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K100257
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberVLOCM0023
Device Catalogue NumberVLOCM0023
Device Lot NumberA3A0158VFY
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2025
Supplement Date Received by Manufacturer05/11/2026
Initial Report FDA Received Date01/23/2025
Supplement Report FDA Received Date05/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient Age55 YR
Patient SexFemale
Patient Weight78 KG
-
-