| Catalog Number |
CAR-02400 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "the doctor found the luer hub/fitting has crack during used on the patient".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The customer returned and one connector assembly, compression fitting, and lidstock analysis.The compression fitting was attached to the connector assembly upon receipt.Signs of use were observed.Visual analysis revealed that both the red arterial luer hub and the blue venous luer hub contained cracks.Stress marks/scratches were present on both luer hubs.Microscopic examination confirmed the damage.The appearance of this damage is consistent with overtightening or repeated tightening of the hubs.No defects or anomalies were observed on any other portion of the returned device.Both luer connections were functionally tested per the product instructions for use (ifu).The ifu provided with this kit instructs the user, "establish and maintain catheter patency.Solution and frequency of flushing a venous access catheter should be established in hospital/institutional policy".A lab inventory syringe filled with water was used to flush each extension line.Leaking was observed from the crack in the red arterial luer hub.A manual tug test confirmed both luer hub were securely attached to their respective lumens.A device history record review was performed, and no relevant findings were identified.The ifu provided with this kit states, "repeated over tightening of bloodlines, syringes and caps will reduce connector life and could lead to potential connector failure." the report of a cracked luer hub was confirmed through complaint investigation of the returned sample.Visual analysis revealed that both luer hubs contained cracks and damage consistent with overtightening or repeated tightening of the hubs.A device history record review was performed, and no relevant findings were identified.Based on the customer report and the sample received, unintentional use error likely caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that "the doctor found the luer hub/fitting has crack during used on the patient".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.The associated emdr report numbers are 9680794-2025-00053 and 9680794-2025-00054.
|
| |
|
Search Alerts/Recalls
|