| Model Number |
MI1250 SYNCHRONY 2 PIN |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Codes |
Head Injury (1879); Failure of Implant (1924)
|
| Date of Event |
01/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user showed a high ground path impedance (gpi) beginning in (b)(6) 2024 and reportedly experienced a trauma at the implant site.The user was re-implanted.
|
| |
|
Event or Problem Description
|
|
The user showed a high ground path impedance (gpi) beginning in aug-2024 after experiencing a trauma at the implant site.The user was reimplanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the reference electrode, likely caused by an external impact, was determined to be the root cause of device failure.A trauma is mentioned in the recipient's report.The investigation results appear to match the problems mentioned in the recipients report.This is a final report.
|
| |
|