| Catalog Number |
367878 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported after using bd vacutainer® sodium heparin (nh) 95 usp units blood collection tubes for karyotyping, erroneous results were seen in 1 sample.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary - bd had not received samples or photos for investigation.Therefore, 100 retention samples from bd inventory were evaluated by visual examination and no issues were observed as all samples met specifications.Complaints for sample quality are under statistical control for the month of december 2024.At this time, further testing is not indicated.Certain assays, in this case karyotyping analysis, are excluded from our protocols.Therefore, we are at the present time, unable to perform internal testing for karyotyping.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode erroneous results.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Bd quality will continue to monitor sample quality complaints.
|
| |
|
Event or Problem Description
|
|
It was reported after using bd vacutainer® sodium heparinn (nh) 95 usp units blood collection tubes for karyotyping, erroneous results were seen in 1 sample.No adverse impact was reported regarding the patient or user.
|
| |
|