|
This supplemental report is being submitted to provide the results of the final investigation.Updated field(s): d4, d8, d9, g1, h2, h3, h5, h6, h8, h11.The device was returned to olympus for inspection, and the reported failure was confirmed.Based on the results of the investigation, olympus confirmed the device wire was broken; however, the most probable cause of the complaint is cause not established likely due to stress of repeated use, external factors, or handling.Olympus will continue to monitor field performance for this device.Should additional relevant information become available, a supplemental report will be submitted.
|