| Model Number |
OTV-SI |
| Medical Device Problem Codes |
Erratic or Intermittent Display (1182); Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the subject device displayed an abnormal image that was flickering.The issue was discovered during reprocessing.There were no reports of patient involvement.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Three attempts were performed to obtain additional information, but no response was received from the customer.Updated fields: d8, d9, h3, h6, h11.The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: defective video connector socket and battery depletion.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: defective video connector socket should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|