| Model Number |
PED2-500-18 |
| Medical Device Problem Codes |
Material Deformation (2976); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id fg15150-0615-1s (lot: 229627677); product id rfx058-115-08 (lot: b797699); product id ped2-500-18 (b721575); medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a neurovascular flow diversion procedure involving a pipeline flex device, the distal end of the device appeared crimped and would not open.The aneurysm was located in the left internal carotid artery at the ophthalmic artery, was unruptured, and had a saccular morphology.The aneurysm had a maximum diameter of 5 millimeters and a neck diameter of 4 millimeters.The distal landing zone was 4.1 millimeters, and the proximal landing zone was 5 millimeters, with normal vessel tortuosity.The physician attempted to deploy the device by unsheathing it in the m1 segment, but the distal end remained crimped and failed to open, although the mid portion did open.The device was resheathed and repositioned at the target location, but the distal end still did not open.This was the third device in a construct, and the decision was made to remove the device entirely.This was the second pipeline flex 5 x 18 device that was used.The first did the exact same thing.It was also removed and discarded.The pipeline was not placed in a bend.More than 50% of the pipeline had been deployed when it failed to open.The pipeline was resheathed less than or equal to 2 times.There were no additional steps or other devices required to open the pipeline.The pipeline was removed from the patient by resheathing and removing with the microcatheter.Dual antiplatelet treatment was administered with a platelet reactivity unit level of 98, and the angiographic result post-procedure was acceptable.The device was resheathed and removed with the microcatheter, and it was never implanted. ancillary devices include: bmx96 guide catheter lot # h00005464 and aristotle 24 guidewire lot # 030518.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a neurovascular flow diversion procedure involving a pipeline flex device, the distal end of the device appeared crimped and would not open.The aneurysm was located in the left internal carotid artery at the ophthalmic artery, was unruptured, and had a saccular morphology.The aneurysm had a maximum diameter of 5 millimeters and a neck diameter of 4 millimeters.The distal landing zone was 4.1 millimeters, and the proximal landing zone was 5 millimeters, with normal vessel tortuosity.The physician attempted to deploy the device by unsheathing it in the m1 segment, but the distal end remained crimped and failed to open, although the mid portion did open.The device was resheathed and repositioned at the target location, but the distal end still did not open.This was the third device in a construct, and the decision was made to remove the device entirely.This was the second pipeline flex 5 x 18 device that was used.The first did the exact same thing.It was also removed and discarded.The pipeline was not placed in a bend.More than 50% of the pipeline had been deployed when it failed to open.The pipeline was resheathed less than or equal to 2 times.There were no additional steps or other devices required to open the pipeline.The pipeline was removed from the patient by resheathing and removing with the microcatheter.Dual antiplatelet treatment was administered with a platelet reactivity unit level of 98, and the angiographic result post-procedure was acceptable.The device was resheathed and removed with the microcatheter, and it was never implanted.Ancillary devices include: bmx96 guide catheter lot # h00005464 and aristotle 24 guidewire lot # 030518.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information received reported after the second 5x18 pipeline did not open, the physician removed the device and ended the case.The patient was left with one 4.25x25 and one 4.5x25 pipeline device.The cause of the failure to open was not determined.
|
| |
|