MAUDE Adverse Event Report: CENTINEL SPINE, LLC PRODISC L SUPERIOR END PLATE MEDIUM 3-STERILE; PROSTHESIS, INTERVERTEBRAL DISC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CENTINEL SPINE, LLC PRODISC L SUPERIOR END PLATE MEDIUM 3-STERILE; PROSTHESIS, INTERVERTEBRAL DISC Back to Search Results
Model Number PDL-M-SP03S
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 12/30/2024
Type of Reportable Event Serious Injury
Event or Problem Description
A patient was implanted with two pdl devices on (b)(6) 2024 at l4/5 and l5/s1, the device at l4/5 subsided less than a week later causing back pain.The surgeon then removed both devices and fused the patient on (b)(6) 2024.
 
Additional Manufacturer Narrative
An mdr is indicated for this complaint.It was reported that a patient was implanted with two pdl devices at l4/5 and l5/s1 on (b)(6) 2024, less than a week later the device at l4/5 subsided causing back pain.The surgeon then removed both devices and fused the patient on (b)(6) 2024.A review of the dhrs found no anomalies associated with the complaint.Complaint trending found that the rate of complaints is within the levels outlined in the risk documentation.A review of the risk documentation found that the hazards associated with the complaint are identified and mitigated to a level where the clinical benefits outweigh the risks.No device evaluation could be completed as the implants were not returned following the removal surgery.Provided imaging confirmed the implant subsidence.The implant removal was completed due to implant subsidence, a reason for the subsidence is unknown.The submission is 1 of 6 devices involved in this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODISC L SUPERIOR END PLATE MEDIUM 3-STERILE
Common Device Name
PROSTHESIS, INTERVERTEBRAL DISC
Manufacturer (Section D)
CENTINEL SPINE, LLC
900 airport road, suite 3b
west chester PA 19380
Manufacturer (Section G)
FRUH VERPACKUNGSTECHNIK AG
allmendstrasse 47
fehraltorf, 8320
SZ   8320
Manufacturer Contact
erin ward
900 airport road
suite 3b
west chester, PA 19380
4848878894
MDR Report Key21230104
Report Number3007494564-2025-00010
Device Sequence Number7578188
Product Code MJO
UDI-Device Identifier00843193114983
UDI-Public(01)00843193114983(17)290731(10)2023-1340
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P050010
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPDL-M-SP03S
Device Catalogue NumberN/A
Device Lot Number2023-1340
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/30/2024
Initial Report FDA Received Date01/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexFemale
-
-