Model Number VLOCM0023
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter during a right knee replacement, when suturing the skin using the continuous suturing technique, the suture broke.It was noted that it lasted five minutes.Another sutures were used for restitching to resolved the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
D10 concomitant products: vlocm0023, vlocm0023 v-loc* 90 4-0 clr 45cm p12x12, (lot #a3g1282vfy) vlocm0023, vlocm0023 v-loc* 90 4-0 clr 45cm p12x12, (lot #a3g1282vfy) vlocm0023, vlocm0023 v-loc* 90 4-0 clr 45cm p12x12, (lot #a3g1282vfy) additional information: b5, d10, g3, h11.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a right knee replacement, when suturing the skin using the continuous suturing technique, four sutures broke.It was noted that it lasted five minutes.Another sutures were used for restitching to resolved the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d3 (manufacturing name, street, manufacturing city), d4 (unique identifier (udi) #), d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the needles were returned attached to the sutures.The sutures were returned broken into multiple fragments.The segments attached to the needles measured 1 inch, 4 1/8 inches and 7 7/8 inches in length.The measurements were taken with an aluminum rule, calibrated asset nh02505.One suture segment was tied in a knot.One foil pouch was returned.Upon inspection, no signs of damage or abnormalities were identified.Functional testing found that the product was returned open and no sealed samples were returned, a tensile strength test could not be performed.The tensile strength of absorbable sutures may be affected by degradation of the suture and/or damage during use after the product has been opened and may impact the validity of any test results.It was reported that sutures broke.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The evaluation detected an unreported condition: off label use.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: progressive loss of tensile strength and eventual absorption of the absorbable sutures occurs by means of hydrolysis.This loss is triggered when exposed to the environment or clinical application.Damage during use may result in inadequate sample length to perform testing and/or deformation of the suture contributing to a loss in tensile strength.Description: the barb and loop end effector design allow for tissue approximation without the need to tie surgical knots.Contraindication: v-loc¿ 90 absorbable wound closure device is not intended to be used by tying surgical knots.Warnings and precautions: the v-loc¿ 90 absorbable wound closure device is intended to be used without the use of anchoring knots to begin or terminate the suture line.Do not tie knots.Tying knots may damage the barbs and potentially reduce their effectiveness.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-LOC 90
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21230758
Report Number1219930-2025-00435
Device Sequence Number7577860
Product Code GAM
UDI-Device Identifier10884521131682
UDI-Public(01)10884521131682(17)260630(10)A3G1282VFY(20)01
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K100257
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model NumberVLOCM0023
Device Catalogue NumberVLOCM0023
Device Lot NumberA3G1282VFY
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2025
Supplement Date Received by Manufacturer01/21/2025
12/12/2025
Initial Report FDA Received Date01/24/2025
Supplement Report FDA Received Date01/29/2025
12/15/2025
Date Device Manufactured07/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient Age62 YR
Patient SexMale
Patient Weight83 KG
-
-