| Model Number |
96280 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that while monitoring patients on a exhibit central station, the central lost all telemetry channels.There was no patient or user harm associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A spacelabs technical support representative walked the user through performing a restart of the exhibit telemetry receiver.Following the restart, the customer confirmed the issue was resolved.Cause of failure was not determined.Note regarding b4, udi field: only the gtin information for the reported device was provided as the serial number was not available.
|
| |
|