| Catalog Number |
AA6114 |
| Medical Device Problem Codes |
Tear, Rip or Hole in Device Packaging (2385); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Urinary Tract Infection (2120)
|
| Date of Event |
01/06/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.Without the benefit of analyzing the device, coloplast cannot comment on the condition of the device.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to the available information, increased hospitalizations for utis in (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
After receiving this complaint, we searched for other complaints and we found none regarding the lot number 9711508.According to the information known, the quality database was checked and revealed the field safety notice associated with the corrective and preventive action z-0844-2025.All the affected products were recalled on the 5 previous years.The investigation resulted in some corrective actions like the change of the packaging foil, new control tools, and retraining of operators.A similar case study was performed based on recall scope.Defect: packaging issue or infection from march 2021 to march 2025: 17 similar cases were found.A medical assessment was also performed which concluded: the risk of infection cannot be eliminated, essentially through a contamination cascade from the outer pouch breach to the external surface of the inner pouch, then to the surgeon¿s glove, then to the patient.Taking into account the multiple risk ratios for each step of the cascade, and based on the sensitivity analysis performed, the final risk is not expected to exceed the usual highest rates reported in literature for each device family.
|
| |
|
Search Alerts/Recalls
|