ETHICON INC. VCL+ UD 8X18IN 0 S/A CT-1 CR; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP840D
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Wound Dehiscence (1154); Purulent Discharge (1812); Unspecified Infection (1930)
Date of Event 12/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.H6 component code: g07002 device not returned to date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure the diagnosis and indication for the index surgical procedure? what was the initial approach for the index surgical procedure? (open, laparoscopic or other)? on what tissue was the suture used? what was the tissue condition (normal, thin, calcified, fragile, diseased)? how was the suture placed (interrupted or continuous)? how was the suture originally tied (multiple knots, square knot, etc.)? please describe the patient manifestations of the reported infection (location, severity, appearance, systemic or local infection).Please provide the onset date/time of infection from the initial surgical procedure.Were there any pre-existing signs/symptoms of active infection prior to this surgical procedure? did the patient receive any prophylactic antibiotics pre-, intra- or post-operation? were cultures performed? if so, please provide the results.How much and what type of drainage is present in this wound? please describe any medical intervention performed including medication name and results.Were any pre-op cleansing procedures changed recently? if yes, please describe.Did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status?.
 
Event or Problem Description
It was reported that a patient underwent a lumbar decompression on (b)(6) 2024 and suture was used.Post-op, the patient experienced a surgical site infection.The patient returned on (b)(6) 2024 to have the infection surgically removed.The patient was doing well as of 30-dec-2024.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: a2, a3a, a4, b7, h6.Additional information was requested, and the following was obtained: please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure: 38 y/o male, 171.4 lbs.24.6 bmi.The diagnosis and indication for the index surgical procedure? lumbar radiculopathy, low back pain.What was the initial approach for the index surgical procedure? (open, laparoscopic or other)? open.On what tissue was the suture used? deep stitch to close.What was the tissue condition (normal, thin, calcified, fragile, diseased)? normal.How was the suture placed (interrupted or continuous)? interrupted.How was the suture originally tied (multiple knots, square knot, etc.)? multiple.Please describe the patient manifestations of the reported infection (location, severity, appearance, systemic or local infection).Surgical site wound dehiscence, purulent drainage.No fever, chills, or other signs of infection on admit to hospital for debridement.Please provide the onset date/time of infection from the initial surgical procedure.Patient states symptoms starting a week post op.Was given oral antibiotics, after round was finished, followed up with surgeon, sent to er for surgery.Were there any pre-existing signs/symptoms of active infection prior to this surgical procedure? no.Did the patient receive any prophylactic antibiotics pre-, intra- or post-operation? preop were cultures performed? if so, please provide the results.Yes.The only culture that came back with any positive result was scant coagulase-negative staphylococcus suggestive of usual skin flora.How much and what type of drainage is present in this wound? report states persistent purulent drainage.Please describe any medical intervention performed including medication name and results.Patient stated a round of oral antibiotics post op.I&d of surgical site, then was sent home from hospital after second surgery antibiotics via picc line.Were any pre-op cleansing procedures changed recently? if yes, please describe.No.Patient was prepped first with isopropyl alcohol, the cleansed with chloraprep twice with 3 minute dry time in between.Did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? none reported.Other relevant patient history/concomitant medications? no.What is the physician's opinion as to the etiology of or contributing factors to this event? none reported.What is the patient's current status? unknown.As of 12/30 was on home iv infusions of antibiotics but otherwise doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 8X18IN 0 S/A CT-1 CR
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN ANGELO
3348 pulliam st
san angelo TX 76905
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key21232733
Report Number2210968-2025-00934
Device Sequence Number20966747
Product Code GAM
UDI-Device Identifier10705031055773
UDI-Public(01)10705031055773(17)290731(10)10309H
Combination Product (Y/N)Y
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberVCP840D
Device Lot Number10309H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2025
Supplement Date Received by Manufacturer01/28/2025
Initial Report FDA Received Date01/24/2025
Supplement Report FDA Received Date01/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
Patient SexMale
Patient Weight78 KG
-
-