| Catalog Number |
VCP840D |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Wound Dehiscence (1154); Unspecified Infection (1930); Impaired Healing (2378); Fluid Discharge (2686)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a lumbar decompression on (b)(6) 2024 and suture was used.Post-op, the patient experienced a surgical site infection.The patient had to return about a month later to have the infection surgically removed.The current condition of the patient was unknown.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.H6 component code: g07002 device not returned to date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure the diagnosis and indication for the index surgical procedure? what was the initial approach for the index surgical procedure? (open, laparoscopic or other)? on what tissue was the suture used? what was the tissue condition (normal, thin, calcified, fragile, diseased)? how was the suture placed (interrupted or continuous)? how was the suture originally tied (multiple knots, square knot, etc.)? please describe the patient manifestations of the reported infection (location, severity, appearance, systemic or local infection).Please provide the onset date/time of infection from the initial surgical procedure.Were there any pre-existing signs/symptoms of active infection prior to this surgical procedure? did the patient receive any prophylactic antibiotics pre-, intra- or post-operation? were cultures performed? if so, please provide the results.How much and what type of drainage is present in this wound? please describe any medical intervention performed including medication name and results.Were any pre-op cleansing procedures changed recently? if yes, please describe.Did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status?.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: a2, a3a, a4, b7, h6.Additional information was requested, and the following was obtained: please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure: 69 y/o female, 128 pounds, 23.4.The diagnosis and indication for the index surgical procedure? lumbar stenosis, low back pain.What was the initial approach for the index surgical procedure? (open, laparoscopic or other)? open.On what tissue was the suture used? deep stitch to close.What was the tissue condition (normal, thin, calcified, fragile, diseased)? normal.How was the suture placed (interrupted or continuous)? interrupted.How was the suture originally tied (multiple knots, square knot, etc.)? multiple please describe the patient manifestations of the reported infection (location, severity, appearance, systemic or local infection).Surgical site ¿unhealing¿ report states seroma vs hematoma, rule out abscess, small opening in incision, no significant fluctuation or drainage, no ither signs or symptoms of infection.Please provide the onset date/time of infection from the initial surgical procedure.Patient states symptoms starting 3-4 weeks post op.Were there any pre-existing signs/symptoms of active infection prior to this surgical procedure? no.Did the patient receive any prophylactic antibiotics pre-, intra- or post-operation? preop.Were cultures performed? yes.The only culture that came back with any positive result was scant coagulase-negative staphylococcus suggestive of usual skin flora how much and what type of drainage is present in this wound? reports states ¿weeping¿ please describe any medical intervention performed including medication name and results.I&d of surgical site.Were any pre-op cleansing procedures changed recently? if yes, please describe.No.Patient was prepped first with isopropyl alcohol, the cleansed with chloraprep twice with 3 minute dry time in between.Did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? none reported.Other relevant patient history/concomitant medications? former smoker, copd, htn, high cholesterol, daily marijuana usage.What is the physician's opinion as to the etiology of or contributing factors to this event? none reported.What is the patient's current status? unknown.As of (b)(6) 2024 reported feeling great.(b)(6) 2025 went to hospital for non healing wound after follow up with pain doctor.
|
| |
|
Search Alerts/Recalls
|
|
|