MAUDE Adverse Event Report: MICROVENTION INC. AZUR CX 18D-LONGER; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION INC. AZUR CX 18D-LONGER; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 45-780308-L
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code Vasoconstriction (2126)
Date of Event 01/03/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was discarded and therefore not available for return and investigation by the manufacturer.However, a video was provided.The investigation is ongoing.Upon completion of the investigation, a supplemental mdr will be submitted.
 
Event or Problem Description
It was reported that during a coil embolization procedure using a emborrhoid technique, the coil was well placed and packed with previous coils, but the pusher was not able to be removed.The coil was attempted to be recaptured, but the coil could not be retracted into the catheter where it became stuck.The coil was attempted to be detached by pulling on it, but the previous coils were also stuck.The physician managed to remove the coil with it remaining outside the catheter and packed.There was no patient injury or intervention reported.Additional information was received that stated there was no intervention required to remove the device.The coil pack kept in place.The case was completed by stopping the intervention after removal of the packed coil due to artery spasm.Patient is ok and will be retreated in a few weeks.
 
Additional Manufacturer Narrative
Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation; however, the customer provided a video of the reported coil device.The video shows the coil implant getting stuck at the distal tip of the catheter when the pusher is being pulled.The retraction attempt is aborted, and the coil implant was pulled out from the distal tip of the catheter.It is then observed that the implant was separated from the pusher.Investigation findings: without the return and physical evaluation of the device, the investigation cannot further examine if a condition existed that would have contributed to the reported event.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Based on a review of the device¿s risk documentation, the reported event did not indicate there were the presence of any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform an analysis of the device, this investigation cannot identify with certainty any potential root causes.Ifu review (additional information can be found in the ifu): potential complications potential complications include, but are not limited to: hematoma at the site of entry, vessel/aneurysm perforation, unintended parent artery occlusion, incomplete filling, vascular thrombosis, hemorrhage, ischemia, vasospasm, edema, coil migration or misplacement, premature or difficult coil detachment, clot formation, revascularization, post-embolization syndrome, and neurological deficits including stroke and possibly death.The physician should be aware of these complications and instruct patients when indicated.Appropriate patient management should be considered.Warnings and precautions ¿ this device is intended for single use only.Do not reuse, reprocess or resterilize.Reuse, reprocessing or resterilization may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness, or death.Reuse, reprocessing, or resterilization may also create a risk of contamination of the device and/or cause patient infection or cross-infection, including, but not limited to, the transmission of infectious disease(s) from one patient to another.Contamination of the device may lead to injury, illness or death of the patient.¿ if a coil must be retrieved from the vasculature after detachment, do not attempt to withdraw the coil with a retrieval device, such as a snare, into the delivery catheter.This could damage the coil and result in device separation.Remove the coil, microcatheter, and any retrieval device from the vasculature simultaneously.Detachment of the coil 34.When the azur detachment controller is properly connected to the delivery pusher, a single audible tone will sound and the light will turn green to signal that it is ready to detach the coil.If the detachment button is not pushed within 30 seconds, the solid green light will slowly flash green.Both flashing green and solid green lights indicate that the device is ready to detach.If the green light does not appear, check to ensure that the connection has been made.If the connection is correct and no green light appears, replace the azur detachment controller.36.Push the detachment button.When the button is pushed, an audible tone will sound and the light will flash green.37.At the end of the detachment cycle, three audible tones will sound and the light will flash yellow three times.This indicates that the detachment cycle is complete.If the coil does not detach during the detachment cycle, leave the azur detachment controller attached to the delivery pusher and attempt another detachment cycle when the light turns green.38.The light will turn red after the number of detachment cycles specified on the azur detachment controller labeling.Do not use the azur detachment controller if the light is red.Discard the azur detachment controller and replace it with a new one when the light is red.39.Verify detachment of the coil by first loosening the rhv valve, then pulling back slowly on the delivery system and verifying that there is no coil movement.If the implant did not detach, do not attempt to detach it more than two additional times.If it does not detach after the third attempt, remove the delivery system.
 
Event or Problem Description
See h11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR CX 18D-LONGER
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICROVENTION INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key21236663
Report Number2032493-2025-00087
Device Sequence Number13030952
Product Code KRD
UDI-Device Identifier00816777024398
UDI-Public(01)00816777024398(11)240905(17)290831(10)0000696970
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123384
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number45-780308-L
Device Lot Number0000696970
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2025
Supplement Date Received by Manufacturer01/30/2025
Initial Report FDA Received Date01/25/2025
Supplement Report FDA Received Date02/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-