|
MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON® SOLERA® VOYAGER¿ SPINAL SYSTEM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED M
|
Back to Search Results |
|
| Model Number |
6550002 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E: first name and last name of initial reporter is unknown.H11: this regulatory report is being submitted due to retrospective review through capa: 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from a user facility (uf) via manufacturer representative regarding spinal products that is used in unknown spinal therapy.It was reported that during inspection, it was found that the drivers were broken while resetting the trays.There is no patient involved in the event and no further complications or symptoms were reported.
|
| |
|
Search Alerts/Recalls
|
|
|