MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON SOLERA 5.5/6.0; SCREWDRIVER Back to Search Results
Model Number 5584109
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a user facility (uf) via manufacturer representative regarding spinal product that will be used in spinal surgery.It was reported that the tip of the driver was broke off.There is no patient involved in the event and no further complications or symptoms were reported.
 
Additional Manufacturer Narrative
E: first name and last name of initial reporter is unknown.H3: product analysis of part# 5584109, lot# ot16f013 visual inspection confirmed the entire torx tip of instrument has been sheared.Optical examination of the fracture surface revealed a fairly flat fracture surface and circular material flow.This type of damage is consistent with torsional overload.H11: this regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SOLERA 5.5/6.0
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key21237042
Report Number1030489-2025-00365
Device Sequence Number18165948
Product Code HXX
UDI-Device Identifier00643169043671
UDI-Public00643169043671
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number5584109
Device Catalogue Number5584109
Device Lot NumberOT16F013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2023
Initial Report FDA Received Date01/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-