• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS STANDOP VOLISTA®; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS STANDOP VOLISTA®; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARDVST229019A
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event site name: (b)(6).Getinge became aware of an issue with one of surgical lights - standop volista.It was stated and confirmed by photographic evidence that rust occurred on the main tube.We decided to report the issue in abundance of caution as any rust particles falling off into sterile field or during procedure may cause contamination.Based on the information collected, it was established that when the event occurred, the surgical lights did not meet its specification, since rust occurrence and/or paint chipping could be considered as technical deficiencies, and in this way the device contributed to event.It is unknown if claimed devices were or were not being used for patient treatment or diagnosis when the event took place as stated by the subject matter expert, peeling paint and/or rust formation were probably caused by: a stagnation of aggressive disinfectant and detergent products due to an inappropriate cleaning protocol.The presence of liquid water or an exposure to humid air due to a high relative humidity of the operating room.The instruction for use indicates the environmental condition for use, the relative humidity must be between 20% and 75%.To prevent any incident the instruction for use mentions to perform daily and monthly inspection in order to detect painting defects, impact marks or other damages.The instruction for use mentions not to clean the device under running water nor spray a solution directly onto the device.Certain cleaning products or procedures may damage the paintwork of the device, which may result in particles falling onto the surgical site during an operation.Fumigation methods are unsuitable for disinfecting the unit and must not be used.The instruction for use mentions to wipe with a dry cloth and to make sure no liquid residue is left on the device after cleaning.To reduce the appearance of rust or the degradation of the surface, the technical manual or maintenance manual mentions in the preventive maintenance protocol to lubricate some parts of device.The instruction for use mentions not to use a damaged device because it may lead to a risk of injury for users or a risk of infection for patients.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Event or Problem Description
Getinge became aware of an issue with one of surgical lights - standop volista®.It was stated and confirmed by photographic evidence that rust occurred on the main tube.We decided to report the issue in abundance of caution as any rust particles falling off into sterile field or during procedure may cause contamination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STANDOP VOLISTA®
Common Device Name
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key21238913
Report Number9710055-2025-0000016
Device Sequence Number12871698
Product Code FSY
UDI-Device Identifier03700712401665
UDI-Public(01)03700712401665
Combination Product (Y/N)N
Initial Reporter CountryEZ
PMA/510(K) Number
K130513
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARDVST229019A
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2025
Initial Report FDA Received Date01/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/26/2020
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-