Catalog Number 204-23-11
Medical Device Problem Codes Crack (1135); Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Synovitis (2094); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 06/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
1038671-2022-01192 d10: 02-012-45-3040 - (b)(6) - logc tibial fit tray cem sz 3f / 4t , 200-02-38 - (b)(6) - three peg patella 38mm, 201-78-82 - (b)(6) - collar fixation pin 2pk 40mm, quick release, 201-78-88 - (b)(6) - 4" drill bit, mod.Hex 2-pk.The product associated with the reported event is within the scope of recall z-0019-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2022-01192.The reason for the revision reported cannot be confirmed from the information provided but may be the result of femoral loosening, prosthesis wear, and cracking.Or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that approximately 70 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, ambulation difficulties, balance problems, discomfort, osteolysis, pain and synovitis.Initial surgery and revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL INSERTS SZ 3, 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key21240409
Report Number1038671-2025-00422
Device Sequence Number20970245
Product Code JWH
UDI-Device Identifier10885862048318
UDI-Public10885862048318
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/12/2018
Device Catalogue Number204-23-11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2022
Initial Report FDA Received Date01/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age62 YR
Patient SexMale
-
-