EXACTECH, INC. PS TIBIAL INSERTS SZ 3, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
204-23-11 |
| Medical Device Problem Codes |
Crack (1135); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Synovitis (2094); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
1038671-2022-01192 d10: 02-012-45-3040 - (b)(6) - logc tibial fit tray cem sz 3f / 4t , 200-02-38 - (b)(6) - three peg patella 38mm, 201-78-82 - (b)(6) - collar fixation pin 2pk 40mm, quick release, 201-78-88 - (b)(6) - 4" drill bit, mod.Hex 2-pk.The product associated with the reported event is within the scope of recall z-0019-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2022-01192.The reason for the revision reported cannot be confirmed from the information provided but may be the result of femoral loosening, prosthesis wear, and cracking.Or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 70 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, ambulation difficulties, balance problems, discomfort, osteolysis, pain and synovitis.Initial surgery and revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.
|
| |
|