Model Number 9-ASD-015
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that on (b)(6) 2025, a 15mm amplatzer septal occluder was chosen for implant using a 8f non-abbott delivery system.During deployment the device, physician noticed that the desired shape was not taken by the device and had a cobra deformation so they have to retrieve the device back.The deformed device was never released from the delivery cable while inside the patient.There was no report of any interaction with cardiac structures during deployment and no report of any angulation/kink noticed with the delivery system.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient was reported stable.
 
Additional Manufacturer Narrative
An event of device deformity was reported.A returned device inspection could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Additionally, a review of the complaint history identified no similar incidents reported from this lot.Based on the information received, the cause of the reported device deformity could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Additional Manufacturer Narrative
An event of device deformity did not confirm via returned device analysis.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Additionally, a review of the complaint history identified no similar incidents reported from this lot.Based on the information received, the cause of the reported device deformity could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.Codes removed: component code: 4755 part/component/sub-assembly term not applicable.Type of investigation: investigation findings: 3221 no findings available.Investigation conclusions: 4315 cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key21240967
Report Number2135147-2025-00401
Device Sequence Number20965674
Product Code MLV
UDI-Device Identifier00811806010120
UDI-Public(01)00811806010120(17)271130(10)8800455
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9-ASD-015
Device Catalogue Number9-ASD-015
Device Lot Number8800455
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/04/2025
Supplement Date Received by Manufacturer02/28/2025
04/07/2025
Initial Report FDA Received Date01/27/2025
Supplement Report FDA Received Date03/11/2025
04/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-