| Model Number |
0283 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited oversensing on the presenting electrogram.The health care professional advised the patient would be seen in clinic for troubleshooting.This rv lead remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited oversensing on the presenting electrogram.The health care professional advised the patient would be seen in clinic for troubleshooting.Additional information from the field representative provided the patient was subsequently seen in clinic.Sensitivity was reprogrammed.No oversensing has been observed following the programming change.The patient continues to be monitored.This rv lead remains in service.No adverse patient effects were reported.
|
| |
|