Model Number 0283
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this right ventricular (rv) lead exhibited oversensing on the presenting electrogram.The health care professional advised the patient would be seen in clinic for troubleshooting.This rv lead remains in service.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead exhibited oversensing on the presenting electrogram.The health care professional advised the patient would be seen in clinic for troubleshooting.Additional information from the field representative provided the patient was subsequently seen in clinic.Sensitivity was reprogrammed.No oversensing has been observed following the programming change.The patient continues to be monitored.This rv lead remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE SG
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112 5798
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21241423
Report Number2124215-2025-03724
Device Sequence Number18266585
Product Code NVY
UDI-Device Identifier00802526431388
UDI-Public00802526431388
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/20/2017
Device Model Number0283
Device Catalogue Number0283
Device Lot Number323139
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2025
Supplement Date Received by Manufacturer02/03/2025
Initial Report FDA Received Date01/27/2025
Supplement Report FDA Received Date02/04/2025
Date Device Manufactured06/20/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-