• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL PROTAPER SHAP.21MM/S1 *NOT CE*; FILE, PULP CANAL, ENDODONTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL PROTAPER SHAP.21MM/S1 *NOT CE*; FILE, PULP CANAL, ENDODONTIC Back to Search Results
Catalog Number A041022110112
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 01/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
 
Event or Problem Description
In this event it is reported that a protaper shap.21mm/s1 *not ce* file broke during use.After treatment patient complained of mild swelling and pain in the root canal.Broken part was found and retrieved.Patient's symptoms disappeared after follow up treatment.
 
Additional Manufacturer Narrative
Summary: involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1753500).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique (sequence followed and motor setting not communicated), overuse (number of uses not communicated), excessive wear, or material issue (no analysis of the broken fragments possible).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key21241921
Report Number8031010-2025-00009
Device Sequence Number12885612
Product Code EKS
UDI-Device IdentifierD716PTURS121
UDI-PublicD716PTURS121
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 02/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA041022110112
Device Lot Number1753500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date01/22/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/22/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/27/2025
Supplement Report FDA Received Date02/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-