Visual inspection: during the investigation, scratches on the outer housing of the valves, was determined.Permeability test: a permeability test has shown that the m.Blue is permeable.Computer controlled test: to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical position.The results show that the m.Blue is not operating within the acceptable tolerance in either position.An accelerated outflow of m.Blue could be determined.Adjustment test: the m.Blue valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valve, deposits were found in m.Blue.Results: based on our investigation results, we can determine an accelerated outflow at m.Blue in both positions, horizontal and vertical.The deposits visible in the valve have led to the change in flow rate.Deposits of natural substances found in the body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of proteins can impair the integrity of the valve.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|