| Model Number |
ENF-P4 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the rhino-laryngo fiber scope had brown foreign material on the bending rubber of the distal end.The issue was found during reprocessing.There was no patient involvement.
|
| |
|
Event or Problem Description
|
|
It was reported that the rhino-laryngo fiber scope had brown foreign material on the bending rubber of the distal end.The issue was found during reprocessing.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: d8, h3, h4, h6, h11 the device was returned to olympus for inspection and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause not established.Olympus will continue to monitor field performance for this device.
|
| |
|