• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. RHINO-LARYNGO FIBERSCOPE; RHINO-LARYNGOFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. RHINO-LARYNGO FIBERSCOPE; RHINO-LARYNGOFIBERSCOPE Back to Search Results
Model Number ENF-P4
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that the rhino-laryngo fiber scope had brown foreign material on the bending rubber of the distal end.The issue was found during reprocessing.There was no patient involvement.
 
Event or Problem Description
It was reported that the rhino-laryngo fiber scope had brown foreign material on the bending rubber of the distal end.The issue was found during reprocessing.There was no patient involvement.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: d8, h3, h4, h6, h11 the device was returned to olympus for inspection and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause not established.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RHINO-LARYNGO FIBERSCOPE
Common Device Name
RHINO-LARYNGOFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key21252848
Report Number3002808148-2025-01469
Device Sequence Number9161135
Product Code EOB
UDI-Device Identifier04953170059032
UDI-Public04953170059032
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K011869
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberENF-P4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/07/2025
Supplement Date Received by Manufacturer04/11/2025
Initial Report FDA Received Date01/28/2025
Supplement Report FDA Received Date04/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-