Model Number 6900PTFX
Medical Device Problem Code Device Stenosis (4066)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 01/03/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported that a patient with a 25mm 6900ptfx mitral valve has been placed under consideration for a valve-in-valve procedure after an implant duration of 5 years, 3 months due to stenosis.The patient presented with heart failure.The tmvr has not yet been scheduled.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated: b4, g3, g6, h2, h6 (component code, type of investigation, investigation findings, investigation conclusions).Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.A definitive root cause is unable to be determined, however, patient likely caused or contributed.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated: b4, b5, b7, g3, g6, h2.H11: corrected data: h6 -impact code.
 
Event or Problem Description
It was reported that a patient with a 25mm 6900ptfx mitral valve underwent a valve-in-valve procedure after an implant duration of 5 years, 6 months due to stenosis.The patient presented with heart failure.The tmvr procedure was performed with a 26mm 9755rsl transcatheter valve.The patient was in recovery post procedure.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated: b4, g3, g6, h2, h6, h11.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.There is insufficient information available regarding patient or procedural factors known to cause or contribute to stenosis.Therefore, a definitive root cause cannot be conclusively determined.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.H11: corrected data: b2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES, LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key21252909
Report Number2015691-2025-00503
Device Sequence Number12822897
Product Code DYE
UDI-Device Identifier00690103163582
UDI-Public(01)00690103163582(17)230309
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/09/2023
Device Model Number6900PTFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2025
Supplement Date Received by Manufacturer02/07/2025
05/30/2025
08/01/2025
Initial Report FDA Received Date01/28/2025
Supplement Report FDA Received Date02/27/2025
06/23/2025
08/01/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Other; Life Threatening;
Patient Age74 YR
Patient SexFemale
-
-