| Model Number |
6900PTFX |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Code |
Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
01/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm 6900ptfx mitral valve has been placed under consideration for a valve-in-valve procedure after an implant duration of 5 years, 3 months due to stenosis.The patient presented with heart failure.The tmvr has not yet been scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b4, g3, g6, h2, h6 (component code, type of investigation, investigation findings, investigation conclusions).Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.A definitive root cause is unable to be determined, however, patient likely caused or contributed.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b4, b5, b7, g3, g6, h2.H11: corrected data: h6 -impact code.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm 6900ptfx mitral valve underwent a valve-in-valve procedure after an implant duration of 5 years, 6 months due to stenosis.The patient presented with heart failure.The tmvr procedure was performed with a 26mm 9755rsl transcatheter valve.The patient was in recovery post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b4, g3, g6, h2, h6, h11.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.There is insufficient information available regarding patient or procedural factors known to cause or contribute to stenosis.Therefore, a definitive root cause cannot be conclusively determined.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.H11: corrected data: b2.
|
| |
|