• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CRVD AGG BLADE 4.2MM 5PK; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CRVD AGG BLADE 4.2MM 5PK; ARTHROSCOPE Back to Search Results
Catalog Number 283412
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
This is report 2 of 4 for (b)(4).It was reported that preoperatively to an unspecified surgical procedure, it was observed that the description on the box of an unknown cutting instrument device did not match the content.According to the report, the customer ordered curved shaver blades.It said on the box that the box contained five curved shaver blades but only one was curved and four were not curved.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown but was noted to have occurred in december 2024.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: d1, d4, g4 (510k), h4: this report is for an unknown cutting instrument.Part and lot number are unknown.Without the specific part number; the expiration date, udi number, 510-k number and device manufacture date are unknown.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
This is report 2 of 4 for (b)(4).It was reported that preoperatively to an unspecified surgical procedure, it was observed that the contents of the box did not match the description of the crvd agg blade 4.2mm 5pk device.According to the report, only one out of the five blades was curved.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown but was noted to have occurred in (b)(6) 2024.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: b5, d1, d4, g4, h4: brand name, common device name, procode, catalog, lot number, udi, expiration date, 510(k), device manufacture date corrected.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: d9, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found the device inside the sealed blister and missing the curved part at the shaft.No other anomalies were noted.The manufacturer performed an evaluation with the following results: final manufacturing step was not completed; the blade should be curved per manufacturing step prior to being packaged.Per the internal operating procedures, 100% of the products are to be inspected following the bending operation.Therefore, it has been concluded that only an operator mistake can lead to this failure.It has been confirmed that all operators were trained to the internal operating procedures.The assignable root cause has been linked to human error and as the production site is no longer operating, awareness cannot be performed in production.The overall complaint was confirmed as the observed condition of the crvd full radius bld 4.2mm 5pk would have contributed to the complained device issue.Based on the information currently available, a product issue was identified during the investigation of the sample received and attributed to manufacturing.This product issue will be addressed through j&j medtech orthopaedics quality system.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRVD AGG BLADE 4.2MM 5PK
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key21252911
Report Number1221934-2025-00389
Device Sequence Number8243715
Product Code HRX
UDI-Device Identifier10886705025923
UDI-Public10886705025923
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K131191
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number283412
Device Lot Number268L592
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2025
Supplement Date Received by Manufacturer01/28/2025
02/25/2025
03/13/2025
Initial Report FDA Received Date01/28/2025
Supplement Report FDA Received Date01/28/2025
02/26/2025
03/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-