BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOG BC WING; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 382633
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
No additional information.
 
Additional Manufacturer Narrative
Investigation results: the complaint of needle retraction issues could not be confirmed from the photographs and videos that were provided for investigation.The photographs showed unit packages and fully retracted needles.The videos showed two examples of the safety mechanism being activated and the needle fully retracting within the specified time.Although the images, videos, and manufacturing records do not support the complainant¿s description of the reported event, the complaint has been documented and will continue to be monitored as part of ongoing efforts to identify potential manufacturing related issues.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.
 
Additional Manufacturer Narrative
This mdr is the result of a complaint record review discovering an additional device that needed capture.H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.E1.Address information was not provided, therefore, xx was used as a place holder.
 
Event or Problem Description
It was reported that bd insyte autog bc wing needle was slow to retract.The following information was provided by the initial reporter: 20 gauge needles - nurses have noticed difficulty in the needles not retracting back in once they have been used.When they push the button to retrack the needle, the needle gets stuck and they have to use their finger to assit it to retract.Even though they push the button multiple times it does not retract.However when they tested a needle that has not been used on a patient, the needle retracts as desinged.This issue may be related to the lot, but it only occurs when used on a patient.They have said they do not have issues on the 22 gauge needles.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE AUTOG BC WING
Common Device Name
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key21254184
Report Number1710034-2025-00080
Device Sequence Number9117069
Product Code FOZ
UDI-Device Identifier00382903826339
UDI-Public(01)00382903826339(10)4241245
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K201075
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number382633
Device Lot Number4241245
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2024
Supplement Date Received by Manufacturer03/07/2025
Initial Report FDA Received Date01/28/2025
Supplement Report FDA Received Date03/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-