| Catalog Number |
382633 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: the complaint of needle retraction issues could not be confirmed from the photographs and videos that were provided for investigation.The photographs showed unit packages and fully retracted needles.The videos showed two examples of the safety mechanism being activated and the needle fully retracting within the specified time.Although the images, videos, and manufacturing records do not support the complainant¿s description of the reported event, the complaint has been documented and will continue to be monitored as part of ongoing efforts to identify potential manufacturing related issues.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr is the result of a complaint record review discovering an additional device that needed capture.H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.E1.Address information was not provided, therefore, xx was used as a place holder.
|
| |
|
Event or Problem Description
|
|
It was reported that bd insyte autog bc wing needle was slow to retract.The following information was provided by the initial reporter: 20 gauge needles - nurses have noticed difficulty in the needles not retracting back in once they have been used.When they push the button to retrack the needle, the needle gets stuck and they have to use their finger to assit it to retract.Even though they push the button multiple times it does not retract.However when they tested a needle that has not been used on a patient, the needle retracts as desinged.This issue may be related to the lot, but it only occurs when used on a patient.They have said they do not have issues on the 22 gauge needles.
|
| |
|
Search Alerts/Recalls
|