INTEGRA LIFESCIENCES SWITZERLAND SAR DIRECTLINK ICP MODULE; DIRECTLINK - MODULE Back to Search Results
Catalog Number 826828
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
2 of 2 reports.Other mfg report number: 3013886523-2025-00018.A facility reported a microsensor basic kit (id: 626631) was placed in the operative room, the patient was transferred to the intensive care unit (icu).The icu staff was never able to retrieve data.The direct link monitor (unknown id) and sensor never displayed a green active light and the connection to the patient monitor was not showing up either.No activity from the sensor via the directlink and patient monitor, therefore, the sensor was explanted.
 
Additional Manufacturer Narrative
Updated fields: d1, d2a, d4, d9, g3, g6, h2, h3, h6, h11.The directlink module (id (b)(6)) was not returned for evaluation and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIRECTLINK ICP MODULE
Common Device Name
DIRECTLINK - MODULE
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key21254401
Report Number3013886523-2025-00019
Device Sequence Number20846856
Product Code GWM
UDI-Device Identifier10381780515784
UDI-Public10381780515784
Combination Product (Y/N)N
PMA/510(K) Number
K152670
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number826828
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2025
Supplement Date Received by Manufacturer07/24/2025
Initial Report FDA Received Date01/28/2025
Supplement Report FDA Received Date08/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
-
-