| Model Number |
SIG30AVM |
| Medical Device Problem Codes |
Misfire (2532); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: egiauxl - egiauxl endogia ultra univ xl stapler, lot# p3l0697.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a roux-en-y procedure, the surgeon was able to press the green button but could not squeeze the handle.The device did not fire.To resolve the issue, a new handle was used with the same reload.It was noted that the procedure time was extended for ten minutes.There was no patient injury.Pli-20: medtronic's initial evaluation of the incident found that the reload was partially fired with the interlock engaged.The jaws were open.Staple pushers were visible at the 2.2cm cut line.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.No information regarding the specific customer issue that occurred with the device was available.Upon examination of the sample, it was found that the device partially fired.Visually, the reload was partially fired with the interlock engaged.The jaws were open and the staple pushers were visible at the 2.2cm cut line.The product analysis noted evidence that the device was not used as intended.The partial fire condition with interlock engagement may occur when the firing handle has not been completely cycled.If the instrument return knobs are retracted at any point once the firing cycle has begun, the reload will engage into safety interlock and prevent further attempts to fire by ceasing the placement of staples and tissue transection and prevent patient harm.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to completely fire the reload will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis and/or leakage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a roux-en-y procedure, the surgeon was able to press the green button but could not squeeze the handle.The device did not fire.To resolve the issue, a new handle was used with the same reload.It was noted that the procedure time was extended for 10 minutes.There was no patient injury.Medtronic's initial evaluation of the incident found that the reload was partially fired with the interlock engaged.The jaws were open.Staple pushers were visible at the 2.2cm cut line.
|
| |
|