| Catalog Number |
CDH23P |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a gastric procedure surgery, after fired, noted the staples malformed.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 1/28/2025 d4 batch # unk attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 2/24/2025.D4 batch #555c18.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the cdh23p device arrived with the anvil missing.The breakaway washer was not present and there were staples present indicating that the device had not been fired.A new washer was placed on a test anvil.The device was tested for functionality and it fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.No malformed staples were observed.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: ensure tissue thickness lays flat and is evenly distributed within the device.Excess tissue on one side may result in unacceptable staple formation and can result in staple line leakage.Do not attempt to manipulate the adjusting knob during the firing sequence.Doing so may result in malformed staples, incomplete cut line, bleeding, and leakage from the staple line and/or difficulty removing the device.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 555c18, and no non-conformances related to the reported complaint condition were identified.
|
| |
|
Search Alerts/Recalls
|
|
|