EXACTECH, INC. TRULIANT PS CEM FEM PS CEM RIGHT SZ 3.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-020-11-0335 |
| Medical Device Problem Codes |
Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Pain (1994); Scar Tissue (2060); Synovitis (2094); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Report number: 1038671-2023-01967, 1038671-2024-04923 and 1038671-2024-04924 d10 concomitants: 204-34-04 - fluted stem extension 40l x 14 mm 4208683 204-70-00 - tibial stem ext.Screw 4163167 multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-01967, 1038671-2024-04923 and 1038671-2024-04924 the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and product information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
As reported via legal documentation the patient had a right knee replacement on 23mar2018.Approximately 4 years and 10 months after the initial procedure the patient had a right knee revision on (b)(6) 2023.Patient has experienced prosthesis wear, femoral loosening, osteolysis and synovitis.Revision op report on (b)(6) 2023 diagnosis: failed polyethylene component, right total knee due to wear with loose femoral component.Findings: there was found to be a clean wound with clear synovial joint fluid.There was evidence of polyethylene wear with encapsulated polyethylene debris within the synovium and a complete synovectomy was performed.There was found to be a loose femoral component from the cement mantle and there was an un-resurfaced patella due to a thin, native, worn patella from previous surgery.The patient was awakened and transferred to recovery in stable condition.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, disassembly and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and product information were not provided.D10 concomitants: 02-022-35-3511 - truliant tib imp ps insert sz 3.5 11mm (b)(6).02-022-45-3535 - truliant tib fit tray cem sz 3.5f / 3.5t (b)(6).204-34-04 - fluted stem extension 40l x 14 mm (b)(6).204-70-00 - tibial stem ext.Screw (b)(6).H6: corrected health effect and medical device clinical codes.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 58 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, severe pain, significant scarring, swelling, effusion, inflammation, knee popping/clunking, knee instability, difficulty walking, antalgic gait, anxiety, and emotional distress.Revision surgery operative notes were provided.The surgeon performed a revision of the right knee polyethylene and femoral component.Patient left the operating room in stable condition.Post operative diagnosis noted that the tibial polyethylene component had disengaged from the tibial tray component.No further issues or complications were reported.No additional information is available.
|
| |
|