• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN? EL ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN? EL ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D141
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) and right ventricular (rv) lead exhibited out of range shock impedance measurements greater than 200 ohms, this was present at the last office interrogation.No adverse patient effects were reported.This device remains in service.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) and right ventricular (rv) lead exhibited out of range shock impedance measurements greater than 200 ohms, this was present at the last office interrogation.No adverse patient effects were reported.This device remains in service.Additional information was received that this icd was explanted approximately two months after this event.This device was received with no additional information.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) and right ventricular (rv) lead exhibited out of range shock impedance measurements greater than 200 ohms, this was present at the last office interrogation.No adverse patient effects were reported.This device remains in service.Additional information was received that this icd was explanted approximately two months after this event.This device was received with no additional information.The 'no problem detected' conclusion code was selected based on analysis of the returned product.Analysis found no evidence of device anomalies outside the bounds of normal medical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN? EL ICD VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21256468
Report Number2124215-2025-04166
Device Sequence Number8038351
Product Code LWS
UDI-Device Identifier00802526534225
UDI-Public(01)00802526534225(17)170217(21)189079
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/17/2017
Device Model NumberD141
Device Catalogue NumberD141
Device Lot Number189079
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2025
Supplement Date Received by Manufacturer03/18/2025
05/22/2025
Initial Report FDA Received Date01/28/2025
Supplement Report FDA Received Date03/19/2025
05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/17/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
Patient SexMale
-
-