• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH FULL RING HOFFMANN LRF Ø180MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH FULL RING HOFFMANN LRF Ø180MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Catalog Number 49332180
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
As reported: ring broke on hlrf frame in patient home.The patient noticed a noise on christmas eve and then attended clinic on new year's eve for review.Decision remains outstanding as to whether or not patient will be revised (adverse consequences).
 
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
As reported: ring broke on hlrf frame in patient home.The patient noticed a noise on christmas eve and then attended clinic on new year's eve for review.Decision remains outstanding as to whether or not patient will be revised (adverse consequences).
 
Additional Manufacturer Narrative
Correction - please refer to b2, h1, d9 - the device was not returned.The complaint couldn't be confirmed, since the device was not returned for evaluation and no other evidence were provided.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Additional Manufacturer Narrative
Correction: please refer to section d9 the device was returned and h6 (method and results codes).The complaint could be confirmed, since the information for evaluation matches the alleged failure.The device inspection revealed the following: visual examination of the construction received has shown that 3 products are broken.One full ring ø180mm and two static strut.Visual examination of the received full ring ø180mm shows that is broken as reported.In addition, it is visible that the breakage section is fixed to no movement in dislocated position.Investigator reviewed the received information together with engineering and noted: unfortunately, we received no information or image of the complete hoffman lrf construct, nor any details on how it was installed on the patient.Therefore, we cannot understand how all the parts were assembled.The assembly with the three broken parts was most likely an unstable construct, and the parts had to withstand a great deal of force, which led to the three fractures.After the fractures, additional stabilizing material was most likely added, such as the two static struts at the ring fracture to stabilize it, and a threaded rod between the broken static struts to further stabilize it.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Although the issue is most likely related to procedure, the root cause could not be conclusively determined.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.If more information is provided, the case will be reassessed.
 
Event or Problem Description
As reported: ring broke on hlrf frame in patient home.The patient noticed a noise on christmas eve and then attended clinic on new year's eve for review.Decision remains outstanding as to whether or not patient will be revised (adverse consequences).01/16/2026 - two additional devices were received upon product return.Both devices are broken as well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL RING HOFFMANN LRF Ø180MM
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key21260502
Report Number0008031020-2025-00226
Device Sequence Number18156515
Product Code KTT
UDI-Device Identifier07613252610625
UDI-Public(01)07613252610625(11)240219(10)6459AA
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K233741
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number49332180
Device Lot Number6459AA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/03/2025
Supplement Date Received by Manufacturer02/19/2025
01/26/2026
Initial Report FDA Received Date01/28/2025
Supplement Report FDA Received Date03/19/2025
02/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-