| Catalog Number |
68-210 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b.5., d6b., h.6.
|
| |
|
Event or Problem Description
|
|
Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed a cracked valve seat diaphragm valve.As per the investigation procedure, creases and wear abrasion were completed and none of the observations were found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported deflation.This record is for the right side.Device was explanted and replaced.
|
| |
|