• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL SDN. BHD. ADJ FLANGE NON LATEX NASA; AIRWAY, NASOPHARYNGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL SDN. BHD. ADJ FLANGE NON LATEX NASA; AIRWAY, NASOPHARYNGEAL Back to Search Results
Catalog Number 18542032
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that: "the item itself is changing color.It should be green but some items have turned completely black, while others you can see them changing from green to black." the issue was identified prior to use.Asscociated complaints 8040412-2025-00002, 8040412-2025-00003, and 8040412-2025-00018.
 
Additional Manufacturer Narrative
Qn# (b)(4).Full udi not available as the correct lot# was not provided by the customer.
 
Additional Manufacturer Narrative
(b)(4).The actual complaint device was not returned; however, the customer provided a photo for evaluation.Based on the review of the photo, the complaint was confirmed.A device history record review was performed and no relevant findings were identified.The manufacturing site reports a capa has been opened to address this issue.
 
Event or Problem Description
It was reported that: "the item itself is changing color.It should be green but some items have turned completely black, while others you can see them changing from green to black." the issue was identified prior to use.Asscociated complaints 8040412-2025-00002, 8040412-2025-00003, 8040412-2025-00018 and 8040412-2025-00014.
 
Event or Problem Description
It was reported that: "the item itself is changing color.It should be green but some items have turned completely black, while others you can see them changing from green to black." the issue was identified prior to use.Associated complaints 8040412-2025-00002, 8040412-2025-00003, 8040412-2025-00018 and 8040412-2025-00014.
 
Additional Manufacturer Narrative
Qn# (b)(4).Lot number of the device identified as kma22a0318.A device history record review was performed on lot# kma22a0318 and no relevant findings were identified.Section d.4.Udi# corrected to (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADJ FLANGE NON LATEX NASA
Common Device Name
AIRWAY, NASOPHARYNGEAL
Manufacturer (Section D)
TELEFLEX MEDICAL SDN. BHD.
perak, west malaysia
Manufacturer (Section G)
TELEFLEX MEDICAL SDN. BHD.
lot no : pt2577 jalen perusahaan
4 kamunting industrial estate
perak, west malaysia 34600
MY   34600
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key21262680
Report Number8040412-2025-00014
Device Sequence Number4336706
Product Code BTQ
UDI-Device Identifier14026704336170
UDI-Public(01)14026704336170(11)220128(17)261228(10)KMA22A0318(21)
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number18542032
Device Lot NumberKMA22A0318
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/02/2025
Supplement Date Received by Manufacturer02/20/2025
03/04/2025
Initial Report FDA Received Date01/29/2025
Supplement Report FDA Received Date02/20/2025
03/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-