PERFUSION SYSTEMS PENDITURE LAA EXCLUSION SYSTEM; LEFT ATRIAL APPENDAGE CLIP, IMPLANTABLE Back to Search Results
Model Number LAAC50
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information that during use of appenditure device, it was reported that the patient had an elevated pulmonary artery (pa) pressure of 70/38 and thickened left atrial appendage (laa).The clip was placed with an inferior approach and immediately moved after release at the u end of the device, the tip end stayed in place.It was recaptured two more times and released with similar results.The device was used to complete the procedure.There was no adverse patient effect associated with this event.Medtronic received additional information that customer did not use a sizer and never had for any appendage devices.
 
Additional Manufacturer Narrative
This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENDITURE LAA EXCLUSION SYSTEM
Common Device Name
LEFT ATRIAL APPENDAGE CLIP, IMPLANTABLE
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key21262971
Report Number2184009-2025-00075
Device Sequence Number9412528
Product Code PZX
UDI-Device Identifier00763000791476
UDI-Public00763000791476
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/22/2024
Device Model NumberLAAC50
Device Catalogue NumberLAAC50
Device Lot NumberT3L025X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2024
Initial Report FDA Received Date01/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
-
-