• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE INTREPID TRANSFORMER; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE INTREPID TRANSFORMER; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065752885
Medical Device Problem Codes Suction Problem (2170); Obstruction of Flow (2423)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that surgeon noticed trouble with the suction while using an ophthalmic irrigation and aspiration transformers.The patient impact and details were not reported.Additional information has been requested but none received till date.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information is provided in sections b.5 and h.6.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
The customer reported that during surgery surgeon noticed something blocking the port and not allowing proper suction while using an ophthalmic irrigation and aspiration transformers.The procedure was completed by opening up new transformer until one was opened that worked properly.No patient harm was reported.
 
Event or Problem Description
The customer reported that during surgery surgeon noticed something blocking the port and not allowing proper suction while using an ophthalmic irrigation and aspiration transformers.The procedure was completed by opening up new transformer until one was opened that worked properly.No patient harm was reported.
 
Additional Manufacturer Narrative
Additional information is provided in sections h.6 and h.11.A sample was not received at the manufacturing site for evaluation for the report of occlusion and suction issue; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Because a sample was not received and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, the root cause for customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.All handpieces are 100% visually inspected prior to being released from the manufacturing facility.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTREPID TRANSFORMER
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21263571
Report Number2523835-2025-00081
Device Sequence Number4276210
Product Code HQC
UDI-Device Identifier00380657528851
UDI-Public(01)00380657528851(17)270930(11)241021(10)173XDY
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K121555
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065752885
Device Lot Number173XDY
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/13/2025
Supplement Date Received by Manufacturer02/04/2025
04/13/2025
Initial Report FDA Received Date01/29/2025
Supplement Report FDA Received Date02/19/2025
04/28/2025
Date Device Manufactured10/21/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-