EXACTECH, INC. ONE PEG PATELLA 32MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
200-03-32 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported via legal documentation the patient had a left knee replacement on (b)(6) 2014.Approximately 4 years and 8 months after the initial procedure the patient had a left knee revision on (b)(6) 2018 due to accelerated and preventable wear.Patient experienced prosthesis wear, instability, loosening, failure of implant, ambulation or postural difficulties, discomfort, osteolysis, joint laxity, swelling/ edema and pain.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Report number: 1038671-2024-00009 d10 concomitants: 2818119 02-010-01-0220 - logic femoral ps cem left sz 2 2819413 02-012-45-2020 - lgc tibial fit tray cem sz 2f / 2t multiple mdr reports were filed for this event, please see associated reports: 1038671-2024-00009.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, loosening, and/or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|