Catalog Number 72203380
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Perforation (2001)
Date of Event 01/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: h2: corrected data on: g2.
 
Event or Problem Description
It was reported that during an arthroscopy, the tip on the healicoil 5.5mm pk anchor broke when screwing into the bone.The broken pieces were removed from inside the patient using a grasper.The procedure was completed using a s+n back up device into an additional drilled bone hole.There was a surgical delay of less than 30 minutes and no further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference number: case- (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALICOIL SA PK 5.5MM W/3 UB-BL CBBL LT
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key21265379
Report Number1219602-2025-00234
Device Sequence Number14587290
Product Code MBI
UDI-Device Identifier00885554026313
UDI-Public885554026313
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K152566
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72203380
Device Lot Number2150354
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2025
Supplement Date Received by Manufacturer01/27/2025
Initial Report FDA Received Date01/29/2025
Supplement Report FDA Received Date01/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age43 YR
Patient SexMale
-
-