ARTHREX, INC. ANGLED REAMER SLEEVE, 20°; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number ANGLED REAMER SLEEVE, 20°
Medical Device Problem Codes Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 01/06/2025, it was reported by a sales representative via sems-06748654 that an ar-9676 angled reamer was stuck inside of the ar-9597-20 angled reamer sleeve.This occurred during a case with no effect on the patient.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.One unpackaged ar-9597-20 angled reamer sleeve batch number: 37622320 was received for investigation.Visual evaluation of the returned device noted damage to the device with superficial scratches observed, as well as striations in the inner diameter.Functional testing was performed by inserting the returned ar-9676 angled reamer drive shaft batch: 022304 to the returned ar-9597-20 angled reamer sleeve and it was noted that the device was met with friction and not able to go into the ar-9597-20 angled reamer sleeve smoothly as intended.The most likely cause for the reported failure can be attributed to misuse due to a force perpendicular to the drive shaft being applied while the device is spinning.This applied force creates an excess of friction and heat that ultimately has the potential to cause the device to seize from functioning as intended.Application of this force perpendicular to the drive shaft is not needed for the device to function, nor is it recommended that the user do this during normal clinical use.When used normally, this issue is unlikely to present itself, which suggests that the handling of the device has a large impact on whether or not it will seize.For this reason, the direct cause is considered to be misuse of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGLED REAMER SLEEVE, 20°
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21265664
Report Number1220246-2025-00109
Device Sequence Number4538589
Product Code LXH
UDI-Device Identifier00888867343450
UDI-Public(01)00888867343450(11)240122(10)37622320
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberANGLED REAMER SLEEVE, 20°
Device Catalogue NumberAR-9597-20
Device Lot Number37622320
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2025
Supplement Date Received by Manufacturer02/27/2025
Initial Report FDA Received Date01/29/2025
Supplement Report FDA Received Date03/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/22/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-