| Model Number |
WA69416M |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the device's jaw was misaligned and had slight bend.The issue was found during inspection for use for a therapeutic laparoscopic total hysterectomy, laparoscopic oophorectomy, cystectomy, bilateral salpingectomy, excision of endometriosis, lysis of adhesions procedures there were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected field: d4.The device was returned to olympus for inspection and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: mechanical overload.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|