• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN PROGRAMMER; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN PROGRAMMER; NO MATCH Back to Search Results
Model Number 3650
Medical Device Problem Code Interrogation Problem (4017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
Related manufacturer reference number: 2017865-2025-11845.It was reported the patient presented for implant procedure.While finalizing the programming on the leadless pacemaker (lp), conductive telemetry with the programmer failed which caused the lp to reprogram itself.The lp was successfully restored.There were no patient consequences.
 
Additional Manufacturer Narrative
The reported complaint of interrogation failure was confirmed.Based on the information provided, a false positive read of the device configuration occurred during the finalization of the vlp.As a result, the programmer inadvertently misconfigured the vlp device as a 2 lp system leading to an interrogation failure due to a data corruption.The root cause was unable to be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Common Device Name
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key21266824
Report Number2017865-2025-11846
Device Sequence Number8533449
Product Code NIK
UDI-Device Identifier05414734500661
UDI-Public(01)05414734500661(10)
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2025
Supplement Date Received by Manufacturer03/07/2025
Initial Report FDA Received Date01/29/2025
Supplement Report FDA Received Date03/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age88 YR
Patient SexFemale
-
-