| Model Number |
3650 |
| Medical Device Problem Code |
Interrogation Problem (4017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Related manufacturer reference number: 2017865-2025-11845.It was reported the patient presented for implant procedure.While finalizing the programming on the leadless pacemaker (lp), conductive telemetry with the programmer failed which caused the lp to reprogram itself.The lp was successfully restored.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint of interrogation failure was confirmed.Based on the information provided, a false positive read of the device configuration occurred during the finalization of the vlp.As a result, the programmer inadvertently misconfigured the vlp device as a 2 lp system leading to an interrogation failure due to a data corruption.The root cause was unable to be determined.
|
| |
|