• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; NASAL CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; NASAL CANNULA Back to Search Results
Model Number OJR414VT
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A distributor in south dakota reported that the ojr414vt optiflow junior 2 ventilator transition kit was found damaged.There was no reported patient involvement.
 
Additional Manufacturer Narrative
(b)(6).Fisher & paykel healthcare (f&p) is in the process of completing our investigation.We will provide a follow-up report upon completion of our investigation.
 
Additional Manufacturer Narrative
(b)(4).Method: the subject ojr414vt optiflow junior 2 ventilator transition kit was not returned to fisher & paykel healthcare (f&p) for evaluation.F&p requested for further information from the healthcare facility, including the sequence of events, photographs, diagrams of the device set-up etc.However, no photographs and limited information was provided.Our investigation is based on the information, and description of events provided by the distributor, previous investigations of similar complaints and our knowledge of the product.Results: the distributor stated that the ojr414vt optiflow junior 2 ventilator transition kit was damaged.There was no reported patient consequences reported by the distributor.Conclusion: without the subject device, we are unable to determine the exact cause of the reported fault.The optiflow junior nasal cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and pediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.All optiflow junior cannula are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula tube joint, as well as the swivel grip joint, if there are any failures the entire batch is put aside for further investigation.The subject ojr414vt optiflow junior 2 ventilator transition kit would have met the required specifications at the time of production.The user instructions which accompany the ojr414vt optiflow junior 2 ventilator transition kit show in pictorial format the correct placement and fitting of the cannula, and provide the following warnings: "appropriate patient monitoring (e.G., oxygen saturation) must be used at all times.Failure to monitor the patient may result in loss of therapy, serious harm or death." "ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient." "do not stretch the cannula on application; this may cause increase pressure to the patient's skin.If necessary, the cannula may be repositioned.".
 
Event or Problem Description
A distributor in south dakota reported that the ojr414vt optiflow junior 2 ventilator transition kit was found damaged.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key21268342
Report Number9611451-2025-00079
Device Sequence Number8505264
Product Code BTT
UDI-Device Identifier09420012448972
UDI-Public(01)09420012448972(10)2103168671(11)240604(17)270604
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
K222197
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOJR414VT
Device Catalogue NumberOJR414VT
Device Lot Number2103168671
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/31/2024
Supplement Date Received by Manufacturer02/04/2025
Initial Report FDA Received Date01/29/2025
Supplement Report FDA Received Date03/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-